ASTM-E1603:2011 Edition
$40.63
E1603/E1603M-11 Standard Practice for Leakage Measurement Using the Mass Spectrometer Leak Detector or Residual Gas Analyzer in the Hood Mode
Published By | Publication Date | Number of Pages |
ASTM | 2011 | 6 |
ASTM E1603/E1603M-11
Historical Standard: Standard Practice for Leakage Measurement Using the Mass Spectrometer Leak Detector or Residual Gas Analyzer in the Hood Mode
ASTM E1603/E1603M
Scope
1.1 This practice covers procedures for testing the sources of gas leaking at the rate of 1 × 10−8 Pa m3/s (1 × 10−9 standard-cm3/s at 0°C) or greater. These test methods may be conducted on any object that can be evacuated and to the other side of which helium or other tracer gas may be applied. The object must be structurally capable of being evacuated to pressures of 0.1 Pa (approximately 10−3 torr).
1.2 Three test methods are described;
1.2.1 Test Method A—For the object under test capable of being evacuated, but having no inherent pumping capability.
1.2.2 Test Method B—For the object under test with integral pumping capability.
1.2.3 Test Method C—For the object under test as in Test Method B, in which the vacuum pumps of the object under test replace those normally used in the leak detector (LD).
1.3 Units—The values stated in either SI or std-cc/sec units are to be regarded separately as standard. The values stated in each system may not be exact equivalents: therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Keywords
helium leak test; helium mass spectrometer leak test; hood leak test; leak testing; mass spectrometer leak test; Leak testing–gas; Mass spectrometer leak detector (MSLD); Helium leak testing; Hood leak test; Residual gas analyzer (RGA)
ICS Code
ICS Number Code 23.040.99 (Other pipeline components)
DOI: 10.1520/E1603_E1603M-11
PDF Catalog
PDF Pages | PDF Title |
---|---|
1 | Scope Referenced Documents Terminology Summary of Practice |
2 | Significance and Use Basis of Application Interferences FIG. 1 FIG. 2 FIG. 3 |
3 | Apparatus Instrument Calibration System Calibration and Test Procedure FIG. 4 |
4 | FIG. 5 |
5 | Apparatus Preparation of Apparatus Instrument Calibration System Calibration and Test Procedure Apparatus Preparation of Apparatus Instrument Calibration System Calibration and Test Procedure Keywords |